51 CHANGI BUSINESS PARK CENTRAL 2 THE SIGNATURE #04-05 SINGAPORE 486066

SALES & ENQUIRIES: +65 86613002

ISO 9001:2015

Atomic Absorption Spectrophotometer

Used for testing the metal element concentration analyze in agriculture, chemical, environmental study, food, mining, and petrochemical, pharmaceutical industry.

Description

Atomic Absorption Spectrophotometer is an integrated flame/graphite furnace atomization system where the graphite system is fully automatic. Flame temperature is continuously adjustable between 2300-2900ºC, which makes it possible to choose the best atomization temperature. Rich Oxygen Flame will not pollute the environment and is not harmful for human bodies. It’s a break-through in flame AAS analysis.

Feature 

  • The changeover of the integrated flame and graphite furnace atomizer is automatically controlled featuring easy operation and time saving which eliminates human labor
  • To perform flame emission analysis, flame emission burner head can be installed for alkali metals as K, Na etc.
  • Provided with automatic 6-lamp turret, automatic adjustment of lamp current and optimization of light beam position
  • Parameters like Wavelength scanning, peak picking, change in spectral bandwidth, optimization of position parameters, automatic ignition and gas flow setting are done automatically
  • The graphite furnace analysis is fully reliable and automatic where it is adopting FUZZY-PID and dual curve mode light-controlled temperature control technique, temperature auto-correction technique, ensures fast heating, good temperature reproducibility and high analytical sensitivity
  • The temperature control accuracy is less than 1%
  • Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact
  • Automatic standard sample preparation, automatic correction of sampling probe depth, automatic tracing and correction of liquid surface height in the sample vessel, with sampling accuracy of 1% and reproducibility of 0.3%
  • Equipped with alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
  • Protection function and alarms for insufficient carrier gas and protective gas pressure, insufficient cooling rate supply and over-heating in graphite furnace system
  • Adopting large-scale programmable logic array and Inter I2C bus technology
  • For long term reliability of the whole electronic system, provided with European type sockets and AMP adapters
  • AAS analysis is made under windows operating system which is easy-to-use with fast parameter setting and optimization
  • In order to increase the sensitivity in flame analysis, two high performance HCLs can be mounted on the lamp turret
  • The D2 lamp and S-H background correction capability at 1A and 1.8A respectively is better than 30 times
  • Provided with detector R928 photo multiplier with high sensitivity and wide spectral range
  • Data processing system: provided with analytical methods – Working curve auto-fitting standard addition method, automatic sensitivity correction, automatic calculation of concentration and content with windows operating system software
  • Experiments can be repeated 1~99 times along with automatic calculation of mean value, standard deviation and relative standard deviation
  • Equipped with multi task functions- sequential determination of multi-elements in the same sample, condition reading- with model function and result printing- measured data and final analytical report printout in excel
  • Provided with standard RS-232 serial port connection

Specification

O2 Enriched Flame Yes
Auto Sampler Yes
Flame Yes
High Performance Lamp Yes
Graphite Furnace Type Yes
Wavelength Range 190-900nm
Wavelength Accuracy Better than ±0.25nm
Resolution Two Spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley peak energy ratio less than 30%
Baseline Stability 0.004A/30min
Lamp Turret Motorized 6-lamp turret
Lamp Current Adjustment
Wide Pulse Current 0~25mA
Narrow Pulse Current 0~10mA
Lamp Power Supply Mode 400Hz Square wave pulse 100Hz Narrow Square wave pulse + 400Hz wide square wave pulse
Optical System
Monochromator Single beam, Czerny-Turner design grating Monochromator
Grating 1800l/mm
Focal Length 277mm
Blazed Wavelength 250nm
Spectral bandwidth 0.1nm, 0.2nm, 0.4nm, 1.2nm Auto Switch Over
Flame Atomizer
Burner 10cm single slot all-titanium burner
Spray Chamber Corrosion resistant all-plastic spray chamber
Nebulizer High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7ml/min
Emission Burner It is Provided
Graphite Furnace
Temperature Range RT~3000ºC
Heating Rate 2000ºC/sec
Graphite Tube Dimensions (L*OD) 28*8 mm
Characteristic Mass Cd≤0.8*10-12g, Cu≤5*10-12g, Mo≤1*10-11g
Precision Cd≤3%, Cu≤3%, Mo≤4%
Graphite Furnace Auto Sampler
Sample Tray Capacity 55 sample vessels and 5 reagent vessels
Vessel Material Polypropylene
Vessel Volume 3ml for sample vessel, 20ml for reagent vessel
Min. Sampling Volume 1µl
Repeatable Sampling Times 1-99 times
Sampling System Accurate dual pump system with 100µl and 1ml injectors
Air-C2H2 Flame Rich-Oxygen Air-C2H2 Flame
Cu: Characteristic Concentration ≤0.025mg/L
Cu: Detection Limit ≤0.006mg/L
Ba: Characteristic Concentration ≤0.22mg/L
Al: Characteristic Concentration ≤0.4mg/L
Main Unit Dimensions 1070 x 490 x 580 mm
Main Unit Weight 140kg
Graphite Furnace Dimensions 420 x 420 x 460 mm
Graphite Furnace Weight 65kg
Auto Sampler Dimensions 400 x 290 x 290 mm
Auto Sampler Weight 15kg

Send An Enquiry

You’re currently viewing Atomic Absorption Spectrophotometer